DISS Console Gas Outlet

DISS Console Gas Outlet

The ACMD gas outlet are designed for delivery of medical gas from a central pipeline system.Suitable for patient room,operating areas,surgery centers and dental offices. This Ohmeda standard ACMD gas outlet provides a full color-coded for easy gas identification.

All outlets are 100% individually hydrostatically tested from maximum gas flow and 100% gas tightness tested to ensure safety and are backed by our industry leading 3 year Warranty.

We also have other configurations available to match your requirements.

  • Features
  • Specifications
  • Gallery
  • Related Products
  • Video

Features

High Quality

  • Modular structure, convenient maintenance

  • 100% air tightness test and hydrostatically tested

  • 10,000 times of plugging and unplugging

  • Available with box and without box 

  • Swivel inlet copper pipe for easy installation

Perfect Connection

  • Connect the flowmeter using the corresponding adapter

  • Installation method: Console type, Wall mounted type, Medical pendant type, Bed head unit type, etc

Safety And Reliable

  • Accepts only DISS gas specific adapters 

  • Complies with NFPA 99 and CGA G-4.1 standard

  • The socket of each gas shall be gas-specific for the attachment of its base

  • Indexed to prevent interchangeability of gas services

Color Standard

  • Full color Code on the cover(NFPA Color Codes & ISO/CSA Color Codes)

  • Color-coded for easy gas service identification

Easy Installation & Maintenance 

  • 360 swivel inlet copper pipe for easy installation

  • Universal rough-in assembly, can be replace at any time

  • Modular structure, convenient maintenance

Warranty

  • 3 Years

Specifications

Technical Details

  • Identify gases using the USA color standard

  • 360°Swivel Inlet Pipe for easy installation,safety system for Non-crossing of Gases

  • Gas types include oxygen, vacuum, air, nitrogen, carbon dioxide, exhaust gas, etc.

  • Universal Rough-In Assembly to accept Quick Disconnects 

Test

  • 100% air tightness test , leak tested and hydrostatically

  • 10,000 times of plugging and unplugging

  • Complies with NFPA 99 and CGA G-4.1 standards DIMENSIONS units in inch (millimeter)

Installation methods

  • Long Copper Tube or threaded interface

  • Console type, Wall mounted type etc.

  • Secondary insertion and removal to prevent unintentional pull-out and personal safety

  • Cover with detachable protection

Physical

  • Packing Size: 460mm*360mm*220mm (Console type)( 22pcs 1 Carton )

  • Weight: G.W: 11KG; N.W: 10KG (Console type)

Material

  • Outlet panel: Zinc Alloy

  • Valve part: Brass

  • Gas pipe: Copper

  • Installing bracket: Stainless steel 

  • Gas adapter: Stainless steel 

  • Panel: Aluminum

Ordering Information

  • A-CON-DS-OXY

  • CON=Installation Type (CON: Console, CONW: Wall mounted)

  • DS= DISS

  • OXY=Gas Type(AIR: Medical Air, N20: Nitrogen, CO2: Carbon Dioxide) 

  • ( Example:A-CON-DS-OXY :Metal panel type console DISS oxygen gas outlets.)

Gallery

IMG_4158 拷贝.jpg
IMG_4159 拷贝.jpg
IMG_4160 拷贝.jpg
_MG_0243-DISS1 拷贝.jpg
_MG_0227 拷贝.jpg
_MG_0228 拷贝.jpg

Video